laboratory equipments

Laboratory Hazardous Waste Storage Container-Small Type

222.jpg
333.jpg
PRODUCT INQUIRY
The Laboratory Hazardous Waste Storage Container adopted standard marine container as the cabinet, which is suitable for temporary storage kinds of toxic and hazardous laboratory wastes; It consists of container, internal storage area, exhaust system of ventilation, explosion-proof power supply system, leakage recovery system, waste gas purification system,intelligent control system, temperature control system, fire extinguishing device, personal protective device, remote (video) monitoring system, cloud monitoring platform, electrostatic protection device etc.
PRODUCT DISPLAY
TECHNICAL PARAMETERS
Model
YKH-40A
Overall dimension
 L 6000mm x W 2440mm x H 2900mm
The ventilation purification system
Explosion-proof frequency and centrifugal,acid mist absorbed.
Cooling system
Explosion-proof frequency conversion air conditioning.
Power supply
2.2KW, AC  220V
Optional accessories
Remote monitoring and alarm system
Standard configuration
1. Explosion-proof air conditioner: 1unit
2. Explosion-proof variable frequency exhaust fan: 1unit
3. Acid mist deodorization purification device: 1unit
4. Explosion-proof temperature and humidity detector: 1pc
5. Fixed combustible gas detector: 1pc
6. Explosion-proof fluorescent lamp: 2pcs
7. Explosion proof emergency light: 1pc
8. Alarm light: 2pcs
9. Positive pressure air breathing apparatus: 1set
10. Fireman' s outfits: 1set
11. Fireman’s shoes: 1set
12. No-power fan: 3units
13. Individual spray cleaning system: 1set
14. Spray cooling system: 1set
15. Leaking fluid recovery pool: 1pc
16.Storage shelves: 6pcs, PP storage cabinets: 7pcs.


PRODUCT STRUCTURE
Ventilation exhaust and waste purification system
The container using frequency conversion anti-corrosion fan for real time ventilation, Generally the exhaust fan ventilate every 10 minutes.When the staff enters the container, the exhaust fan can be adjusted to a strong ventilation mode, and the toxic and harmful gas will be purified and exhausted.
Internal storage for hazardous waste
PP cabinet storage and shelves storage
Safety and anti-explosion system
In view of the flammable and explosive characteristics of laboratory waste, all electrical products are anti-explosion to ensure safe storage.
Fluid Leakage Recovery System
The inner wall adopts the anti-corrosion coating, Through the special installation method, the leaking liquid natural collection. No need special leak collecting tank,Ensure the internal leakage of waste is reliably recycled into the leak collection pool. Meanwhile, the inner wall of the system is protected from corrosion in harsh environment.
Fire emergency System
The container achieved A-level fire standards. It equipped with special fire cabinets,fire extinguishers, emergency escape door exits and emergency indicators,  which can ensure fire emergency and escape in case of sudden disaster.
Personal protective system
The containers are equipped with personal protective cabinets, positive pressure air breathing device and personal protective clothing, protective shoes and gloves. Which can ensure ensure the safety of personnel in the event of sudden disaster. the outdoor also equipped with spray purification devices which can clean and purify the workers in time after exposure to hazardous materials
Automatic monitoring alarm and emergency disposal system
The container installed temperature, humidity and combustible gas detection alarm system, the external alarm system will emit sound and light alarm if the relevant indicators exceeded . Meanwhile the exhaust fan automatically adjusts to the strong ventilation mode, then purify and exhaust toxic gas.
Anti-static system
The container equipped with anti-static column and anti-static floor to completely discharge the static hazard
Temperature control system
The container using double-layer insulation board, it also equipped with temperature control system which consists of temperature detection device, explosion-proof air-conditioning and sprinkler system,The explosion-proof air-conditioning will automatic work once the temperature exceeded . At the same time, Meanwhile you can also automatically or manually activated sprinkler system to cooling in case of abnormal high temperature or fire.

USE SUPPORT
1、Site requirements:Site dimension 18m×6m,  The bottom of the equipment must be higher than the highest water level, The site should be stayed away from the living area and the crowd, Should be outside the area of flammable, explosive, other dangerous goods warehouses and high-voltage transmission line protection area.
2、Hardened and leak-proof base
3、Leaking fluid collecting pool
4、Shower sink;
5、Protective fences;
6、Sun Shelter;
7、Poisonous and harmful warning signs;
8、water and power supply;

Related Products

Laboratory Hazardous Wast...

CONTACT
info@yonglekang.com
dc@yonglekang.com
+86-731-89723840
+86-731-84225947
Buliding B8, Huanchuangyuan, No.2450 Yuelu West Avenue,High-tech Development Zone, Changsha, Hunan, China
2018-2020 CHANGSHA YONG LE KANG CO.,LTD. ALL RIGHT RESERVED.